The 5240 TECSource is a popular temperature controller for lower power applications. With 28W of TEC control and support for thermistor sensors (10µA or 100µA bias current), it is the ideal choice when higher power or broader sensor support is not needed. It also has an auxiliary DC fan power supply for controlling external fans, such as those in our 207 LaserMounts. Housed in a compact enclosure, the 5240 is an excellent mix of features, size, and price.
AutoTune Automatic PID Calculation
All Arroyo Instruments temperature controllers feature AutoTune for automatic PID calculation. With AutoTune, you no longer need to fiddle with PID parameters to find values that work with your application. Simply hook up your mount, set the appropriate limits, and start the AutoTune process. The instrument will automatically calculate PID parameters that will work with your mount.
Excellent Stability
The 5240 offers excellent short-term and long-term stability. At less than 0.004°C stability over one hour, and 0.01°C over 24 hours, the 5240 provides an excellent solution for precision temperature control.
Fully Adjustable PID
The 5240 comes with eight factory set gain settings for the temperature control loop, which are sufficient for most applications, but also offers fully adjustable PID settings, allowing you to fine-tune the response of the instrument. Also, because the loop is digital and uses the temperature (not sensor resistance) as the control variable, variations in sensor sensitivity across temperature (primarily seen with thermistors) does not affect the unit as it does with analog-based PID loops.
Integrated Fan Supply
The 5240 includes an adjustable 4.5V to 9V DC power supply for power external fans, such as those on our 240 and 260 Series LaserMounts. By incorporating the power supply, it eliminates the need for an external power supply and simplifies cabling to the mount. If you will be using this power supply, make sure you choose the 1260B or 1261B TECSource Cable, which includes the additional wiring for the power supply.
Standard Computer Interfaces
The 5240 includes USB computer interface, allowing for quick and easy connection to a PC for remote operation. In addition, the TECSource’s command set is compatible with ILX and Newport controllers, allowing you to leverage any existing software you may have already developed.
Simple User Interface
The 5240’s user interface is remarkably simple… so easy to use, you’ll have it up and running in no time. Unlike other instruments which rely on 7-segment LED displays and a confusing array of indicators and buttons, the 5240 instead presents the information on a high contrast VFD display in an easy-to-read format. The 5240 displays the temperature setpoint, actual temperature, current, and voltage simultaneously… no need to toggle through the readings like on most instruments. If you need a larger display for viewing at a distance, you can press the knob to toggle to a larger display of the actual temperature. Settings and error messages are in clear English, not cryptic codes and flashing status lights.
Key Specs
Power (A, V, W) | ±4A, ±7V, 28W |
Stability (1 hour, °C) | 0.004 |
Stability (24 hours, °C) | 0.01 |
Temperature
Range (°C) | -99 to 250 |
Resolution (°C) | 0.01 |
Thermistor (100µA) Accuracy at 25°C (°C) | 0.03 |
Thermistor, 100µA Range
Accuracy (±[% reading + kΩ]) | 0.05 + 0.005 |
Range (kΩ) | 0.05 – 45 |
Resolution (kΩ) | 0.001 |
Thermistor, 10µA Range
Accuracy (±[% reading + kΩ]) | 0.05 + 0.05 |
Range (kΩ) | 0.1 – 450 |
Resolution (kΩ) | 0.01 |
Current
Range (A) | ±4 |
Compliance Voltage (V) | >7 |
Max Power (W) | 28 |
Resolution (A) | 0.01 |
Accuracy (±[% value + A]) | 0 + 0.03 |
Noise/Ripple (A, rms) | < 0.003 |
Current Limit Accuracy (A) | 0.05 |
Voltage (measurement only)
Accuracy (±[% reading + V]) | 0 + 0.05 |
Range (V) | ±7 |
Resolution (V) | 0.01 |
Other
TEC Connector | DB-15, female |
External Fan Support | Yes, 4 – 8.5V, 350mA |
Display Type | 2×20 VFD |
Computer Interface | USB 2.0 Full Speed (USB Type B) |
Power | 100V / 120V / 230V 50/60 Hz |
Size (H x W x D) [inches (mm)] | 1.82 (47) x 8.5 (215) x 11.13 (283) |
Operating Temperature | +10°C to +40°C |
Storage Temperature | -20°C to +60°C |
Datasheet | 5240 TECSource Datasheet |
Manuals | 5240 TECSource User’s Manual |
Manuals | Computer Interfacing Manual |
Software | Arroyo Control |
Software | USB Drivers |
Software | LabVIEW Drivers |
Model | 5240 TECSource CAD Model |